俞小娟#, 于传飞#, 张荣建, 武刚, 崔永霏, 郭璐韵, 王兰*. HPLC-FMA法测定单抗药物中聚山梨酯80含量的方法学联合验证J. 药学学报, 2021,56(8): 2276-2281. doi: 10.16438/j.0513-4870.2021-0151
引用本文: 俞小娟#, 于传飞#, 张荣建, 武刚, 崔永霏, 郭璐韵, 王兰*. HPLC-FMA法测定单抗药物中聚山梨酯80含量的方法学联合验证J. 药学学报, 2021,56(8): 2276-2281. doi: 10.16438/j.0513-4870.2021-0151
YU Xiao-juan#, YU Chuan-fei#, ZHANG Rong-jian, WU Gang, CUI Yong-fei, GUO Lu-yun, WANG Lan*. Interlaboratory method validation of HPLC-FMA for determination of polysorbate 80 in monoclonal antibodiesJ. Acta Pharmaceutica Sinica, 2021,56(8): 2276-2281. doi: 10.16438/j.0513-4870.2021-0151
Citation: YU Xiao-juan#, YU Chuan-fei#, ZHANG Rong-jian, WU Gang, CUI Yong-fei, GUO Lu-yun, WANG Lan*. Interlaboratory method validation of HPLC-FMA for determination of polysorbate 80 in monoclonal antibodiesJ. Acta Pharmaceutica Sinica, 2021,56(8): 2276-2281. doi: 10.16438/j.0513-4870.2021-0151

HPLC-FMA法测定单抗药物中聚山梨酯80含量的方法学联合验证

Interlaboratory method validation of HPLC-FMA for determination of polysorbate 80 in monoclonal antibodies

  • 摘要: 对HPLC-FMA (high performance liquid chromatography-fluorescence micelle assay)法测定单抗药物中聚山梨酯80含量的方法开展实验室间的联合验证,以研究该方法在多个实验室间的可应用性和可转移性以及纳入现行版药典的可行性。本文使用两个厂家(J.T.Baker公司和南京威尔)来源的聚山梨酯80在7个实验室开展验证。结果显示,在蛋白浓度≤ 20 mg·mL-1,聚山梨酯80含量在0.05~0.5 mg·mL-1内时,该方法专属性良好;加标样品的准确性分别为92.20%~117.70%(J.T.Baker公司),93.90%~117.20%(南京威尔);组内精密度分别为RSD < 4.30%(J.T.Baker公司),RSD < 2.60%(南京威尔);总体精密度分别为RSD < 5.45%(J.T.Baker公司),RSD < 6.70%(南京威尔);线性相关系数分别为r > 0.98(J.T.Baker公司),r > 0.99(南京威尔)。联合验证结果证明,HPLC-FMA法准确性、精密度、线性和专属性均良好,可应用于不同实验室单抗药物中聚山梨酯80含量的放行检测及稳定性分析。

     

    Abstract: The high performance liquid chromatography-fluorescence micelle assay (HPLC-FMA) method for the content determination of polysorbate 80 in monoclonal antibody drugs was validated to study its applicability and transferability between various laboratories, and the feasibility to be included in the Chinese Pharmacopoeia. Both J.T. Baker and Nanjing Well-sourced polysorbate 80 was used in the collaborative validation of polysorbate 80 content analysis in seven different laboratories. The results show that when the protein concentration was no more than 20 mg·mL-1 and the concentration of polysorbate 80 ranged from 0.05 to 0.5 mg·mL-1, the method had good specificity. The recovery rates of the spiked samples ranged from 92.20% to 117.70% for J.T.Baker and from 93.90% to 117.20% for Nanjing Well. The intra-laboratory precision (%RSD) was less than 4.30% for J.T. Baker, and less than 2.60% for Nanjing Well, while the overall precision was less than 5.45% for J.T. Baker, and less than 6.70% for Nanjing Well. The linear correlation coefficient was more than 0.98 for J.T. Baker and more than 0.99 for Nanjing Well. The results of the collaborative validation prove that the HPLC-FMA method has good accuracy, precision, linearity, and specificity, and could be used for release control analysis of polysorbate 80 content in monoclonal antibodies across different laboratories.

     

/