张驰, 董善年. 反相高效液相色谱法测定人血浆中阿昔洛韦浓度J. 药学学报, 1993, 28(8): 629-632.
引用本文: 张驰, 董善年. 反相高效液相色谱法测定人血浆中阿昔洛韦浓度J. 药学学报, 1993, 28(8): 629-632.
C Zhang, SN Dong. DETERMINATION OF ACYCLOVIR IN HUMAN PLASMA BY RP-HPLCJ. Acta Pharmaceutica Sinica, 1993, 28(8): 629-632.
Citation: C Zhang, SN Dong. DETERMINATION OF ACYCLOVIR IN HUMAN PLASMA BY RP-HPLCJ. Acta Pharmaceutica Sinica, 1993, 28(8): 629-632.

反相高效液相色谱法测定人血浆中阿昔洛韦浓度

DETERMINATION OF ACYCLOVIR IN HUMAN PLASMA BY RP-HPLC

  • Abstract: A RP-HPLC method was developed for the determination of aeyclovir (ACV) in human plasma. ACV was extracted from plasma using ODS, obtained from Sep-Pak C18 cartridge (Waters associates) and reinstalled in a self-made glass tube suitable for extraction of solutes from small volume plasma samples. After extraction, and rinsing the ODS extract with water to remove hydrophilic impurities, ACV was recovered with methanol and chromatographed on YWG C18H37 column, and detected at 254 mm. The mobile phase was 5.0% (v/v) methanol/water. The average recovery of ACV was 73. 5%; the minimum detection concentration of ACV in plasma was 20 ng/ml with a plasma volume of 0. 2 ml and S/N value of 2. A good linear relationship between the peak area ratios and ACV concentrations was found at the ACV concentrations ranging from 0. 2 to 1.0 μg/ml, CACV=0. 840 AACV/Ais+0.002, r=0.9969.

     

/

返回文章
返回