Electrochemical behavior of daunorubicin at Co/GC ion implantation modified electrode and the application of the electrode
-
Abstract
AimTo study the electrochemical behavior of daunorubicin at Co/GC ion implantation modified electrode. MethodsWith Co/GC ion implantation modified electrode as working electrode, daunorubicin was determined by voltammetry in 0.05 mol·L-1 Na2HPO4-KH2PO4(pH 6.82) solution. ResultsA sensitive reductive peak of daunorubicin was obtained by linear sweep voltammetry. The peak potential was -0.60 V (vs SCE)。The peak current was proportional to the concentration of daunorubicin over the range of 2.84×10-8-1.42×10-6 mol·L-1 and 1.42×10-6-1.28×10-5 mol·L-1 with the detection limit of 1.42×10-8 mol·L-1。The reduction wave was applied to the determination of daunorubicin. The electrochemical behavior and reaction mechanism were studied by linear sweep and cyclic voltammetry. ConclusionThe reduction process was quasi-reversible with adsorption characteristics.
-
-