YUAN Yi-Sheng, XING Xiang-Fei, ZENG Ping , ZHOU Xia, . DETERMINATION OF PROPRANOLOL IN HUMAN PLASMA BY REVERSED PHASE HIGH PERFORMANCE LIQIUD CHROMATOGRAPHYJ. Acta Pharmaceutica Sinica, 1987, 22(3): 238-240.
Citation: YUAN Yi-Sheng, XING Xiang-Fei, ZENG Ping , ZHOU Xia, . DETERMINATION OF PROPRANOLOL IN HUMAN PLASMA BY REVERSED PHASE HIGH PERFORMANCE LIQIUD CHROMATOGRAPHYJ. Acta Pharmaceutica Sinica, 1987, 22(3): 238-240.

DETERMINATION OF PROPRANOLOL IN HUMAN PLASMA BY REVERSED PHASE HIGH PERFORMANCE LIQIUD CHROMATOGRAPHY

  • This paper investigates the determination method of propranolol in human plasma by HPLC. YWG-C18H37 column was used with CH3OH-H2O-0.2mol/L KH2PO4 0.2mol/L Na2HPO4(460:40:0.7:4.7)as mobile phaseThe preparation of plasma sample: Transfer 0.5 ml plasma into a 5 ml polystyrene test tube with cap, add 0.8 ml of 0.2% sodium hydroxide solution into the plasma. Propranolol was extracted with ether. External standard was used for quantitation.The assay was evaluated using the reproducibility within day and day-to day, and demonstrated high precision. Their average coefficients of variation were 5.8% and 6.8% respectively. The standard curve was linear (r=0.9999)over the range of 0.05~2.00μg/ml, the average recovery was 95.9±4.5%.