Simultaneous determination of pseudoephedrine and diphenhydramine in human plasma by liquid chromatographytandem mass spectrometry
-
Abstract
AimTo develop and validate a liquid chromatographic/tandem mass spectrometric (LC/MS/MS) method for the simultaneous quantitation of pseudoephedrine and diphenhydramine in human plasma after oral administration of compound preparation. MethodsSamples spiked with the analytes and the internal standard, chlorphenamine, were processed using liquidliquid extraction. The extract was chromatographed on a Diamonsil C18 column. The mobile phase consisted of methanolwaterformic acid (70∶30∶1), at a flow rate of 065 mL·min-1. A Finnigan TSQ tandem mass spectrometer equipped with atmospheric pressure chemical ionization source was used as detector and was operated in the positive ion mode. Selected reaction monitoring (SRM) using the precursor/product ion combinations of m/z 166→148 and m/z 256→167 were used to quantify pseudoephedrine and diphenhydramine, respectively. ResultsThe linear concentration ranges of the calibration curves for pseudoephedrine and diphenhydramine were both 2.0-400.0 μg·L-1. The lower limit of quantification of both analytes was 2.0 ng·mL-1. Each sample was chromatographed within 3.9 min. ConclusionThe method was successfully used in simultaneous determination of pseudoephedrine and diphenhydramine in human plasma after oral administration of a compound preparation containing 60 mg pseudoephedrine chloride and 50 mg diphenhydramine chloride. The method was shown to be suitable for clinical investigation of pseudoephedrine and diphenhydramine pharmacokinetics.
-
-